Abstract :
A standard measuring system (setup) has been developed at VNIIM to be used for the calibration of magnetic flux density gradient (MFDG) coils, sensors and magnetic gradiometers (MG). This measurement standard is based on the calculable quartz coil used for the reproduction of gradient which is uniform in a specified working space. It also includes measuring equipment to compare coils, gradiometers and differential measuring transducers. Theoretical and experimental aspects of the problem are discussed. The developed standard system allows calibrating MFDG coils with an uncertainty of 0.1 % within the range from 0.005 to 0.5 T/m.
Keywords :
calibration; fluxgate magnetometers; magnetic flux; measurement standards; MFDG coils; VNIIM; differential measuring transducers; magnetic flux density gradient; magnetic gradiometers calibration; measurement standard; measuring equipment; quartz coil; sensors; standard measuring system; Calibration; Coils; Density measurement; Extraterrestrial measurements; Magnetic flux; Magnetic flux density; Magnetic sensors; Measurement standards; Sensor systems; Standards development;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements (CPEM), 2010 Conference on
Conference_Location :
Daejeon
Print_ISBN :
978-1-4244-6795-2
Type :
conf
DOI :
10.1109/CPEM.2010.5543530
Filename :
5543530
Link To Document :
بازگشت