DocumentCode
3464253
Title
Ac/dc magnetic flux density standard systems at KRISS
Author
Park, Po Gyu ; Kim, Young Gyun ; Wan-Seop Kim ; Shifrin, V. Ya
Author_Institution
Korea Res. Inst. of Stand. & Sci. (KRISS), Daejeon, South Korea
fYear
2010
fDate
13-18 June 2010
Firstpage
312
Lastpage
313
Abstract
The ac/dc magnetic-flux-density (MFD) standard systems were developed for the calibration of magnetometers and the test of sensors and materials. The dc MFD is maintained to cover the range from 20 μT to 2.5 T with the expanded uncertainties (k = 2) from 4 μT/T to 80 μT/T. The expanded uncertainty (k = 2) of ac MFD standard is 0.2 % in the range of 1 Hz to 20 kHz. We had been participated in international key comparison to achieve the equality and the mutual agreement between standard institutes for the results of calibrations and tests.
Keywords
calibration; magnetic flux; magnetometers; sensors; AC/DC magnetic flux density standard systems; KRISS; MFD standard system; expanded uncertainties; frequency 1 Hz to 20 kHz; magnetic flux density 20 muT to 2.5 T; magnetometers calibration; sensors; Calibration; Magnetic flux; Magnetic flux density; Magnetic materials; Magnetic sensors; Magnetometers; Materials testing; Standards development; System testing; Uncertainty;