Abstract :
MEMS technology has been used more and more in biomedical application for neural prosthetic implantation [1, 2]. These devices, however, will have to endure harsh and corrosive body fluids [3]. Therefore, biostable and hermetic-like packaging is needed to protect the implant. This work reports the package reliability of parylene-based retinal implant using active (i.e., with electrical signals applied) and accelerated lifetime soaking test in saline. Commercial amplifier chips, dummy conduction chips, and discrete components are tested and the failure modes are examined. It is found that the proposed parylene-metal-parylene flexible composite sandwich layers indeed have longer lifetime than traditional inflexible silicone-parylene combination. In addition, the chip size effect on lifetime is observed that smaller chips have longer lifetime under the same protection.
Keywords :
amplifiers; bioMEMS; biological fluid dynamics; biomedical materials; eye; lab-on-a-chip; microfabrication; packaging; polymers; prosthetics; silicones; MEMS technology; accelerated lifetime soaking test; amplifier chips; biomedical application; chip size effect; corrosive body fluids; discrete component; dummy conduction chip; electrical signal; failure mode; hermetic-like packaging; implant protection; neural prosthetic implantation; package reliability; parylene-based flexible retinal implant; parylene-metal-parylene flexible composite sandwich layers; traditional inflexible silicone-parylene combination; Coatings; Corrosion; Implants; Metals; Packaging; Retina; Substrates; Lifetime; MEMS; Packaging; Parylene; Retinal implant;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Solid-State Sensors, Actuators and Microsystems (TRANSDUCERS & EUROSENSORS XXVII), 2013 Transducers & Eurosensors XXVII: The 17th International Conference on
Conference_Location :
Barcelona
Type :
conf
DOI :
10.1109/Transducers.2013.6627341
Filename :
6627341
Link To Document :
بازگشت