• DocumentCode
    3464719
  • Title

    A powerless optical microsensor for monitoring intraocular pressure with keratoprostheses

  • Author

    Ghannad-Rezaie, Mostafa ; Gulari, Mayurachat Ning ; de Melo Franco, R. ; Mian, S.I. ; Chronis, Nikos

  • Author_Institution
    Dept. of Biomed. Eng., Univ. of Michigan, Ann Arbor, MI, USA
  • fYear
    2013
  • fDate
    16-20 June 2013
  • Firstpage
    2708
  • Lastpage
    2711
  • Abstract
    Frequent and accurate measurement of intraocular pressure (IOP) is required in monitoring patients who are at high risk of developing glaucoma. We present a novel electronic-free, wireless, optomechanical implantable IOP microsensor. The microsensor has a total volume of less than 1mm3 and is integrated into a keratoprosthesis (KPro). We performed ex-vivo experiments by implanting the microsensor/KPro assembly into human eyebank globes for two weeks. Our microsensor had 2 mmHg accuracy and less than 1 mmHg zero-drift.
  • Keywords
    bio-optics; eye; medical disorders; microsensors; optical sensors; pressure measurement; pressure sensors; prosthetics; KERATOPROSTHESES; electronic-free wireless optomechanical implantable IOP microsensor; glaucoma; human eyebank globes; intraocular pressure monitoring; powerless optical microsensor; Cornea; Microsensors; Optical device fabrication; Optical fibers; Optical filters; Stimulated emission; Wireless pressure sensor; biocompatible; implantable; intraocular pressure; near infrared;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Solid-State Sensors, Actuators and Microsystems (TRANSDUCERS & EUROSENSORS XXVII), 2013 Transducers & Eurosensors XXVII: The 17th International Conference on
  • Conference_Location
    Barcelona
  • Type

    conf

  • DOI
    10.1109/Transducers.2013.6627365
  • Filename
    6627365