DocumentCode :
3467217
Title :
Active ventilation of above- and below-knee prosthetic devices
Author :
Masi, J. ; Nelson, T. ; Skewes, E.
Author_Institution :
Dept. of Electr. & Biomed. Eng., Western New England Coll., Springfield, MA, USA
fYear :
1990
fDate :
26-27 Mar 1990
Firstpage :
25
Lastpage :
26
Abstract :
In an attempt to overcome the problems of lingering moisture and odor, a ventilated socket, coupled with an active cooling/sensing moisture-removal system, was designed. This yields a prosthesis which is lightweight, is low noise, and has low power consumption, while keeping the added cost at a minimum by using standard materials and components. The design, construction, and test data are presented and discussed
Keywords :
artificial limbs; above-knee prosthetic devices; active ventilation; below-knee prosthetic devices; cooling; lingering moisture; moisture removal; odor; sensing; ventilated socket; Cooling; Fabrication; Leg; Moisture; Prosthetics; Skin; Sockets; Temperature; Testing; Ventilation;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Bioengineering Conference, 1990., Proceedings of the 1990 Sixteenth Annual Northeast
Conference_Location :
State College, PA
Type :
conf
DOI :
10.1109/NEBC.1990.66275
Filename :
66275
Link To Document :
بازگشت