• DocumentCode
    3468467
  • Title

    Verification and uncertainty evaluation of an ac shunt calibration system at power frequencies

  • Author

    Kon, Saytaro ; Yamada, Tatsuji ; Tadokoro, Takuya

  • Author_Institution
    Nat. Metrol. Inst. of Japan (NMIJ), AIST, Tsukuba, Japan
  • fYear
    2010
  • fDate
    13-18 June 2010
  • Firstpage
    343
  • Lastpage
    344
  • Abstract
    An ac shunt calibration system has been developed at NMIJ. The calibration results were discussed with dc calibration results and the expanded uncertainties (k=2) of the system have been estimated to be less than 3.4 μ Ω/Ω for ac resistance and 8.6 μrad for phase angle.
  • Keywords
    calibration; NMIJ; ac shunt calibration system; power frequencies; verification-uncertainty evaluation; Amplifiers; Calibration; Carbon capture and storage; Current transformers; Electrical resistance measurement; Frequency; Phase estimation; Resistors; Shunt (electrical); Uncertainty;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements (CPEM), 2010 Conference on
  • Conference_Location
    Daejeon
  • Print_ISBN
    978-1-4244-6795-2
  • Type

    conf

  • DOI
    10.1109/CPEM.2010.5543797
  • Filename
    5543797