Title :
A microwave frequency standard based on trapped, buffer gas-cooled 171Yb+ ions-some accuracy issues
Author :
Fisk, Peter T H ; Sellars, Matthew J. ; Lawn, Malcolm A. ; Coles, Colin
Author_Institution :
Nat. Meas. Lab., CSIRO, Sydney, NSW, Australia
Abstract :
The performance, as characterised by the Allan deviation, of the trapped 171Yb+ ion frequency standard NML-IT-2 is equal to or better than σy(τ)=6×10-14 τ-1/2, for averaging times less than 1000 s. There is evidence that the measured performance is limited both in the short and medium (τ>1000 s) term by the stability of the hydrogen maser used as a frequency reference. A preliminary investigation of the dependence of the observed clock frequency corrected for the calculated second order Doppler shift on the trap conditions has yielded an approximate set of confidence limits on the calculated values of the second order Doppler shift
Keywords :
frequency measurement; 1000 s; 12.6 GHz; Allan deviation; NML-IT-2; Yb; buffer gas-cooling; clock frequency; confidence limits; frequency reference; hydrogen maser; measured performance; microwave frequency standard; second order Doppler shift; stability; trap conditions; trapped 171Yb+ ions;