    Comparison of temperature coefficient of standard inductor by measuring change in inductance and resistance

  • Author

    Ansari, M.A. ; Jyotsana ; Saxena, Anil Kishore

  • Author_Institution
    Nat. Phys. Lab., Council of Sci. & Ind. Res., New Delhi, India
  • fYear
    2010
  • fDate
    13-18 June 2010
  • Firstpage
    378
  • Lastpage
    379
  • Abstract
    Standard inductors manufactured by General Radio (GR) are widely used by NMIs. The temperature coefficient for these standard inductors is 30 ± 5 ppm/°C, as per manufacturer. NPL India (NPLI) had earlier determined the temperature coefficient of standard inductors by measuring change in inductance. Now by measuring change in resistance, temperature coefficient of inductors is re-determined. Temperature coefficients determined from the two methods are reported here.
  • Keywords
    inductors; General Radio; inductance change measurement; resistance change measurement; standard inductor; temperature coefficient; Electrical resistance measurement; Electromagnetic measurements; Inductance measurement; Inductors; Measurement standards; Temperature;