    A novel method and the instrumentation developed for analog measuring of pulse width (T10 or T50) and rise time (T10-90) of cardiac defibrillator discharges are presented. The input pulse is applied to an analog delay network and to a peak detector which determines reference voltages. Those phase-shifted signals go to a comparator that generates the control pulse. A counter and a decoder excite the LCD display that shows the measured parameters with accuracy better than 0.33%
  • Keywords
    comparators (circuits); defibrillators; peak detectors; prosthetics; pulse measurement; sample and hold circuits; LCD display; LV pulses; analog delay network; analog measuring method; analog memory; cardiac defibrillator discharges; comparator; control pulse; decoder; defibrillating shocks; instrumentation; peak detector; phase-shifted signals; pulse width; reference voltages; rise time; Counting circuits; Decoding; Detectors; Instruments; Pulse generation; Pulse measurements; Signal generators; Space vector pulse width modulation; Time measurement; Voltage;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    [Engineering in Medicine and Biology, 1999. 21st Annual Conference and the 1999 Annual Fall Meetring of the Biomedical Engineering Society] BMES/EMBS Conference, 1999. Proceedings of the First Joint
  • Conference_Location
    Atlanta, GA
  • ISSN
    1094-687X
  • Print_ISBN
    0-7803-5674-8
  • Type

    conf

  • DOI
    10.1109/IEMBS.1999.802395
  • Filename
    802395