NEMA NU 2-2000 performance measurements on an ADAC MCD camera
Author :
Sossi, V. ; Pointon, B. ; Boudoux, C. ; Cohen, P. ; Hudkins, K. ; Jivan, S. ; Nitzek, K. ; deRosario, J. ; Stevens, C. ; Ruth, T.J.
Author_Institution :
TRIUMF, British Columbia Univ., Vancouver, BC, Canada
Volume :
3
fYear :
2000
fDate :
2000
Abstract :
The clinically oriented NEMA-NU2 performance standards have been performed on an ADAC MCD dual head camera. At 1 cm radius the transverse resolution FWHM (FWTM) was measured to be 5.2 (10.0) mm and the axial was 5.0 (10.6) mm. At 10 cm radius the transverse radial resolution was 4.7 (7.4) mm and the transverse tangential was 8.5 (20.9) mm. The axial resolution was 14.0 (41.1) mm. The peak true count rate was 4365 counts/sec at a concentration of 4.3 kBq/ml. The peak noise equivalent count rate (NECR) was 1617 count/sec at 2.9 kBq/ml. The system scatter fraction was 48%. The maximum relative count rate error observed at the NECR peak was -42%. The sensitivity was 1014 cts/sec/MBq in the centre and 867 cts/sec/MBq at an offset of 10 cm. The contrast for the hot spheres ranged from 18.5 (11.6)% to 11.6 (7.9)% for the contrast ratio of 8:1 (4:1). For the cold spheres it ranged from 41% to 25% with an overall background variability of 7%. The residual error in scatter and attenuation correction was approximately 50%. All measurements were performed at the maximum allowed crystal-to-crystal distance of 810 mm
Keywords :
biomedical equipment; cameras; image resolution; positron emission tomography; 1 cm; 10 cm; 810 mm; ADAC MCD camera; NEMA NU 2-2000 performance measurements; PET; attenuation correction; axial resolution; clinically oriented NEMA-NU2 performance standards; cold spheres; maximum allowed crystal-to-crystal distance; medical diagnostic imaging; medical instrumentation; nuclear medicine; overall background variability; residual error; transverse radial resolution; transverse resolution FWHM; Cameras; Electric variables measurement; Electromagnetic scattering; Image quality; Magnetic heads; Measurement standards; Particle scattering; Performance evaluation; Positron emission tomography; Testing;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2000 IEEE
Conference_Location :
Lyon
ISSN :
1082-3654
Print_ISBN :
0-7803-6503-8
Type :
conf
DOI :
10.1109/NSSMIC.2000.949172
Filename :
949172
Link To Document :
بازگشت