Title :
Cuff configuration for reduction of interfering signals in nerve cuff recordings
Author :
Andreasen, Lotte N S ; Struijk, Johannes J.
Author_Institution :
Center for Sensory Motor Interaction, Aalborg Univ., Denmark
Abstract :
A new cuff configuration for reduction of interfering signals in nerve cuff recordings is introduced. It is based on the traditional tripolar configuration, but has additional end-electrodes which are short circuited and floating. The additional end-electrodes reduced the interfering signal by a factor of 4 to 1, and when platinized by a factor of 13 to 1
Keywords :
bioelectric potentials; biomedical electrodes; interference (signal); neurophysiology; Pt; cuff configuration; end-electrodes; interfering signals reduction; nerve cuff recordings; short circuited floating electrodes; traditional tripolar configuration; Biomedical measurements; Circuits; Electrodes; Electromyography; In vitro; Performance evaluation; Samarium; Signal to noise ratio; Silicon compounds; Wire;
Conference_Titel :
[Engineering in Medicine and Biology, 1999. 21st Annual Conference and the 1999 Annual Fall Meetring of the Biomedical Engineering Society] BMES/EMBS Conference, 1999. Proceedings of the First Joint
Conference_Location :
Atlanta, GA
Print_ISBN :
0-7803-5674-8
DOI :
10.1109/IEMBS.1999.802555