DocumentCode
3472473
Title
Noncontact heart rate measurement using a 24 GHz Doppler radar
Author
Wee Ser ; Jufeng Yu ; Xufeng Guo ; Jianmin Zhang ; Ong, Marcus Eng Hock
Author_Institution
VALENS Res. Centre, Nanyang Technol. Univ., Singapore, Singapore
fYear
2013
fDate
9-11 Dec. 2013
Firstpage
1
Lastpage
3
Abstract
Current heart rate measurement methods include the use of ECG or pulse oximeter systems. While these systems are effective, they require skin contact. It has been demonstrated that Doppler radars can be used to measure heart rates too. This paper presents the design of a modified 24 GHz Doppler system. The modified signal conditioning board uses a 2-stage amplifier and a simple dc offset compensation circuit. The design removes the unstable problem and reduces circuit noise. The signal processing algorithm used is also presented. Measurements on 16 volunteers were made using the modified system and promising results have been obtained.
Keywords
Doppler radar; biomedical measurement; cardiology; compensation; Doppler radar; ECG; circuit noise; dc offset compensation circuit; frequency 24 GHz; noncontact heart rate measurement; pulse oximeter systems; skin contact; Biomedical measurement; Doppler radar; Heart rate; Microwave FET integrated circuits; Microwave theory and techniques; Radio frequency; Doppler radar; RF technology; heart rate measurements; radar sensing; remote monitoring;
fLanguage
English
Publisher
ieee
Conference_Titel
Microwave Workshop Series on RF and Wireless Technologies for Biomedical and Healthcare Applications (IMWS-BIO), 2013 IEEE MTT-S International
Conference_Location
Singapore
Type
conf
DOI
10.1109/IMWS-BIO.2013.6756191