Abstract :
This paper describes the continuous development of NMC´s microwave power sensor calibration system using the direct comparison transfer method. The system uncertainties are evaluated for the complex reflection coefficient inputs in terms of real and imaginary components and its magnitude and phase format following the internationally recommended guidelines. Monte Carlo method is also used to numerically check the distribution of the output quantities. The measurement results and the uncertainties in different methods are presented and discussed.
Keywords :
Monte Carlo methods; calibration; sensors; Monte Carlo method; complex reflection coefficient; direct comparison transfer method; microwave power sensor calibration system; Calibration; Coaxial components; Electromagnetic measurements; Microwave sensors; Sensor systems;