DocumentCode :
3473625
Title :
Rebinning errors in coincidence imaging due to depth of interaction
Author :
Ordonez, Caesar E. ; Matthews, Kenneth L., II ; Chang, Wei
Author_Institution :
Rush-Presbyterian St. Luke´´s Med. Center, Chicago, IL, USA
Volume :
3
fYear :
2000
fDate :
2000
Abstract :
In positron emission tomography, uncertainties in-or lack of knowledge of-depths of interaction lead to errors in the positioning of coincidence events. This paper examines the dependence of these positioning errors on depth of interaction. Here, the authors derive analytic expressions that relate rebinning errors to the uncertainties in the measurement of interaction positions including depth of interaction. The results confirm the intuitive notion that rebinning errors are minimized when depth of interaction is measured at least as well as the transverse position
Keywords :
errors; positron emission tomography; PET; analytic expressions; interaction positions measurement; medical diagnostic imaging; nuclear medicine; transverse position; Biomedical imaging; Error analysis; Gamma ray detection; Gamma ray detectors; Geometry; Image analysis; Laboratories; Measurement uncertainty; Position measurement; Positron emission tomography;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2000 IEEE
Conference_Location :
Lyon
ISSN :
1082-3654
Print_ISBN :
0-7803-6503-8
Type :
conf
DOI :
10.1109/NSSMIC.2000.949324
Filename :
949324
Link To Document :