DC performance of a high accuracy thermal transfer standard
Author :
Early, M.D. ; Garcocz, M. ; Christian, L.A.
Author_Institution :
Meas. Stand. Lab. of New Zealand (MSL), Lower Hutt, New Zealand
Abstract :
The dc performance of a widely-used high-accuracy commercial thermal transfer standard is found to be significantly affected by a variable contact resistance on its lower impedance ranges. This has a direct impact on the AC voltages measured by the instrument on these ranges. Cleaning the contacts and ensuring clearance around an input common mode choke effectively resolves the problem.
Keywords :
cleaning; contact resistance; digital voltmeters; inductors; temperature measurement; AC voltages; calibrated precision DC DVM; contact cleaning; high-accuracy commercial thermal transfer standard; input common mode choke; variable contact resistance; Calibration; Cleaning; Electrical resistance measurement; Electromagnetic measurements; Frequency; Impedance; Instruments; Laboratories; Measurement standards; Voltage;