A new indirect calibration method for equivalent source mismatch of a power splitter using measurements at only one-port
Author
Suzuki, K.
Author_Institution
Meas. Stand. Center, Agilent Technol. Int. Japan, Ltd., Hachioji, Japan
fYear
2010
fDate
13-18 June 2010
Firstpage
696
Lastpage
697
Abstract
A new calibration method is described for the equivalent source mismatch (ΓG) of a power splitter. It uses one-port measurements by an ordinary one-port calibration at the test port of a vector network analyzer, with the remaining ports loaded with termination, imperfect load, and a series of three characterized standards, i.e., open, short and load. It also takes advantage of the reciprocity characteristic of the power splitter. As the system impedance is normalized to the impedance of the termination, the termination works as a perfect load in measurements and calculations for the S-parameters of the power splitter.
Keywords
S-parameters; calibration; network analysers; power dividers; S-parameters; equivalent source mismatch; indirect calibration method; one-port calibration; one-port measurements; power splitter; system impedance; vector network analyzer; Calibration; Electromagnetic measurements; Impedance measurement; Measurement standards; Power measurement; Reflection; Scattering parameters; Stability; Testing;