    A new indirect calibration method for equivalent source mismatch of a power splitter using measurements at only one-port

  • Author

    Suzuki, K.

  • Author_Institution
    Meas. Stand. Center, Agilent Technol. Int. Japan, Ltd., Hachioji, Japan
  • fYear
    2010
  • fDate
    13-18 June 2010
  • Firstpage
    696
  • Lastpage
    697
  • Abstract
    A new calibration method is described for the equivalent source mismatch (ΓG) of a power splitter. It uses one-port measurements by an ordinary one-port calibration at the test port of a vector network analyzer, with the remaining ports loaded with termination, imperfect load, and a series of three characterized standards, i.e., open, short and load. It also takes advantage of the reciprocity characteristic of the power splitter. As the system impedance is normalized to the impedance of the termination, the termination works as a perfect load in measurements and calculations for the S-parameters of the power splitter.
  • Keywords
    S-parameters; calibration; network analysers; power dividers; S-parameters; equivalent source mismatch; indirect calibration method; one-port calibration; one-port measurements; power splitter; system impedance; vector network analyzer; Calibration; Electromagnetic measurements; Impedance measurement; Measurement standards; Power measurement; Reflection; Scattering parameters; Stability; Testing;