Power sensor linearity calibration with an unknown attenuator
Author :
Yhland, K. ; Stenarson, J. ; Wingqvist, C.
Author_Institution :
SP Tech. Res. Inst. of Sweden, Boras, Sweden
fYear :
2010
fDate :
13-18 June 2010
Firstpage :
769
Lastpage :
770
Abstract :
This paper describes the linearity calibration of power sensors using the setup for the direct comparison method. To achieve sufficient dynamic range an attenuator is used. The S-parameters of the attenuator need not be known because the effective attenuation can be calculated from two sets of measurements on the device under test. This way the linearity is traced to a monitoring power sensor instead of the S-parameters of the attenuator. The method is verified with a bilateral comparison.
Keywords :
S-parameters; attenuators; calibration; microwave detectors; S-parameters; attenuator; bilateral comparison; direct comparison method; power sensor linearity calibration; Attenuation measurement; Attenuators; Calibration; Dynamic range; Linearity; Microwave devices; Monitoring; Scattering parameters; Sensor phenomena and characterization; Testing;