• DocumentCode
    3478400
  • Title

    A multi-step approach for accurate permittivity measurements of liquids using a transmission line method

  • Author

    Zhao, D. ; Rietveld, G. ; Teunisse, G.M.

  • Author_Institution
    Dutch Metrol. Inst., VSL, Delft, Netherlands
  • fYear
    2010
  • fDate
    13-18 June 2010
  • Firstpage
    767
  • Lastpage
    768
  • Abstract
    High-precision permittivity measurements of liquids are essential for specifying the specific absorption rate (SAR) of radio frequency power into human tissue. At VSL, a commercial unit is used for these measurements, but for reaching a low uncertainly some inherent disadvantages need to be overcome. In this paper, a multi-step approach is proposed using a combination of curve fitting, nonlinear progressive technique and an improved transmission line model. This approach is applicable even without a priori knowledge of the liquid properties. Using this approach, the applicable frequency range of the commercial unit is successfully extended from 200 MHz 3 GHz to at least 100 MHz-4 GHz, and the accuracy is improved from 2 % to 0.2 % 0.3 %.
  • Keywords
    biological techniques; curve fitting; permittivity measurement; transmission lines; SAR; VSL; curve fitting; frequency 100 MHz to 4 GHz; human tissue; liquids permittivity measurement; nonlinear progressive technique; radio frequency power; specific absorption rate; transmission line method; transmission line model; Dielectric liquids; Dielectric measurements; Electromagnetic measurements; Equations; Frequency measurement; Humans; Iterative algorithms; Permittivity measurement; Radio frequency; Slabs;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements (CPEM), 2010 Conference on
  • Conference_Location
    Daejeon
  • Print_ISBN
    978-1-4244-6795-2
  • Type

    conf

  • DOI
    10.1109/CPEM.2010.5544368
  • Filename
    5544368