DocumentCode
3478590
Title
A Neurostimulator Design for Long-term Animal Experiments
Author
Zhou, Jingai ; Woo, Se Joon ; Park, Se Ik ; Lee, Seung Woo ; Chung, Hum ; Kim, Sung June
Author_Institution
Sch. of Electr. Eng. & Comput. Sci., Seoul Nat. Univ., Seoul
fYear
2007
fDate
11-13 Oct. 2007
Firstpage
453
Lastpage
457
Abstract
This article reports on a neural prosthesis stimulation system for long-term use in animal electrical stimulation experiments. The presented system consists of an implantable stimulator which provides continuous electrical stimulation, and an external component which provides preset stimulation parameters and power to the implanted stimulator via a paired RF (radio frequency) coil. A rechargeable internal battery and a parameter memory component were introduced to the implanted neural stimulator. As a result, the external component was not necessary during the stimulation cycles. The implantable stimulator was implemented with IC chips and the electronics, except for the stimulation electrodes, were hermetically packaged in a biocompatible metal case. A polyimide-based gold electrode array was used for realization of the animal implantation test using retinal prosthesis approach.
Keywords
bioelectric potentials; biomedical electrodes; gold; neuromuscular stimulation; prosthetics; Au; IC chips; animal electrical stimulation; biocompatible metal case; implantable stimulator; neural prosthesis stimulation system; neurostimulator design; polyimide-based gold electrode array; retinal prosthesis; Animals; Batteries; Coils; Electrical stimulation; Electrodes; Electronics packaging; Hermetic seals; Integrated circuit packaging; Prosthetics; Radio frequency;
fLanguage
English
Publisher
ieee
Conference_Titel
Frontiers in the Convergence of Bioscience and Information Technologies, 2007. FBIT 2007
Conference_Location
Jeju City
Print_ISBN
978-0-7695-2999-8
Type
conf
DOI
10.1109/FBIT.2007.67
Filename
4524148