Title :
Phase aberration correction using dynamic time warping
Author :
Brenner, A.R. ; Eck, K. ; Engelhard, G. ; Noll, T.G.
Author_Institution :
Dept. of Electr. Eng. & Comput. Syst., Tech. Hochschule Aachen, Germany
Abstract :
Dynamic time warping is an efficient method in speech recognition. In this presentation, it is shown that dynamic time warping can be applied to the estimation of phase aberration in ultrasonic imaging as well. Together with a weighting function that considers the local strength of the inspected ultrasonic signals and a modified partial array reference correction, dynamic time warping proved to be a robust estimation scheme for phase aberration. This is demonstrated for aberrated synthetic and experimental data in a comparative evaluation of dynamic time warping together with well-known aberration correction techniques like cross-correlation and speckle look-back
Keywords :
aberrations; acoustic signal processing; biomedical ultrasonics; medical image processing; phase estimation; ultrasonic imaging; ultrasonic transducer arrays; aberrated synthetic data; cross-correlation; dynamic time warping; inspected ultrasonic signals; local strength; modified partial array reference correction; phase aberration correction; robust estimation scheme; speckle look-back; tissue; ultrasonic imaging; weighting function; Acoustic imaging; Acoustic transducers; Biomedical imaging; Biomedical transducers; Phase estimation; Phased arrays; Speckle; Speech recognition; Ultrasonic imaging; Ultrasonic transducers;
Conference_Titel :
Ultrasonics Symposium, 1995. Proceedings., 1995 IEEE
Conference_Location :
Seattle, WA
Print_ISBN :
0-7803-2940-6
DOI :
10.1109/ULTSYM.1995.495809