• DocumentCode
    3486428
  • Title

    Comparison of experimental techniques for evaluating the correction factor of a rectangular waveguide microcalorimeter

  • Author

    Cui, Xiaohai ; Crowley, T.P.

  • Author_Institution
    Nat. Inst. of Stand. & Technol., Boulder, CO, USA
  • fYear
    2010
  • fDate
    13-18 June 2010
  • Firstpage
    551
  • Lastpage
    552
  • Abstract
    We have evaluated the correction factor of a WR-15 (50 to 75 GHz) rectangular waveguide microcalorimeter with four different techniques. The four methods are in agreement. Our initial uncertainty analysis indicates that the technique with the lowest uncertainty is one that utilized two identical microcalorimeter core sections whose reference planes were connected to each other. This initial uncertainty is smaller than expected, which will require us to re-evaluate uncertainty contributions that were neglected.
  • Keywords
    calorimeters; measurement uncertainty; millimetre wave measurement; rectangular waveguides; WR-15 rectangular waveguide microcalorimeter; correction factor evaluation; frequency 50 GHz to 75 GHz; microcalorimeter core section; uncertainty analysis; Coaxial components; Electromagnetic heating; Electromagnetic measurements; Electromagnetic waveguides; Measurement standards; NIST; Performance evaluation; Power measurement; Radio frequency; Rectangular waveguides;