DocumentCode
3486842
Title
Study on absolute calibration methods for mixer-based NVNA
Author
Zhang, Yichi ; Lin, Maoliu ; Zhang, Zhe
Author_Institution
Sch. of Electron. & Inf. Eng., Harbin Inst. of Technol., Harbin, China
fYear
2010
fDate
13-18 June 2010
Firstpage
585
Lastpage
586
Abstract
This paper studies on the calibration of a 0.1-20 GHz mixer-based two-port coaxial Nonlinear Vector Network Analyzer (NVNA), especially the absolute calibration techniques. And from the 8-term error model, we present two alternative two-port calibrating procedures, whose outcomes are revealed equivalent to those of the traditional “unknown thru” method by experiments. The new methods not only relax the requirements of “thru” line, provide more choices and more convenience for calibration, but also offer a way to examine the rationality and validity of the 8-term error model for NVNA.
Keywords
UHF mixers; calibration; microwave mixers; network analysers; 8-term error model; absolute calibration methods; frequency 0.1 GHz to 20 GHz; mixer-based NVNA; mixer-based two-port coaxial nonlinear vector network analyzer; two-port calibrating procedures; unknown thru method; Calibration; Coaxial components; Computer networks; Distortion measurement; Electromagnetic measurements; Load modeling; Nonlinear equations; Phase measurement; Power measurement; Testing;
fLanguage
English
Publisher
ieee
Conference_Titel
Precision Electromagnetic Measurements (CPEM), 2010 Conference on
Conference_Location
Daejeon
Print_ISBN
978-1-4244-6795-2
Type
conf
DOI
10.1109/CPEM.2010.5544818
Filename
5544818
Link To Document