Title : 
Calibration of Rogowski coils at high currents
         
        
            Author : 
Draxler, K. ; Styblikova, R.
         
        
            Author_Institution : 
Fac. of Electr. Eng., Czech Tech. Univ. in Prague, Prague, Czech Republic
         
        
        
        
        
        
            Abstract : 
This paper describes the calibration of a Rogowski coil (RC) with an integrator assigned for current measurement in the range up to 25 kA at 50 Hz mains frequency. A current loop with 10 turns with a current of 2.5 kA is used for calibration. The magnetic field generated in such a way differs from the field of one turn placed in the centre and passed with a current of 25 kA. The influence of non-homogeneities of winding and the cross-section of the RC may cause a calibration error when using a current loop. The difference in the results between using a current loop and one turn and the measurement uncertainties are given below.
         
        
            Keywords : 
calibration; coils; measurement uncertainty; RC cross section; RC winding nonhomogeneity; Rogowski coil calibration; calibration error; current 2.5 kA; current 25 kA; current loop; current measurement; frequency 50 Hz; integrator; magnetic field; measurement uncertainty; Calibration; Coils;
         
        
        
        
            Conference_Titel : 
Precision Electromagnetic Measurements (CPEM), 2010 Conference on
         
        
            Conference_Location : 
Daejeon
         
        
            Print_ISBN : 
978-1-4244-6795-2
         
        
        
            DOI : 
10.1109/CPEM.2010.5544839