DocumentCode
3487257
Title
Measurement of percent carboxyhemoglobin with pulse-oximetry technique
Author
Lee, Jason C S ; Cheung, Peter W. ; Marble, David R. ; Kenny, Margaret A. ; Landicho, Del
Author_Institution
Microsensor Res. Lab., Washington Technol. Center, Seattle, WA, USA
fYear
1988
fDate
4-7 Nov. 1988
Firstpage
1781
Abstract
Computer simulations based on the photon diffusion theory are carried out to analyze the feasibility of measuring the percent carboxyhemoglobin (%HbCO) using the pulse-oximetry technique. A series of in vitro experiments using a mechanical pulser to simulate the pumping effect of the heart is performed. Both the theoretical and the experimental data show that a pulse-oximetry technique using multiple wavelengths can be applied for the measurement of %HbCO.<>
Keywords
biomedical measurement; blood; proteins; computer simulation; heart pumping effect; mechanical pulser; percent carboxyhaemoglobin measurement; photon diffusion theory; pulse oximetry;