Title :
The Checking of Temperature Stability of Magnetically Hard Material by Means Teslameter of Type H1-8
Author :
Arhipov, A.V. ; Glotov, E.P. ; Darmaev, A.N.
Author_Institution :
FSUE R&P Corp., Moscow
Abstract :
For magnetic measurements serial teslameter type H1-8 in a range (100divide110) mT1 that has an error no more plusmn0,15 % is used. For the operative temperature stability checking in a range of temperatures (18divide100) degC of magnets to be used in microwave devices the new method and additional device were developed. The magnets from alloy CS-25 of different manufactures have disorder of values of reversible temperature coefficient of an induction from - 1,9 % up to - 4,8 % on 100degC as has been shown. The effect of samples form on measurements is not shown
Keywords :
magnetic materials; magnetic variables measurement; microwave devices; magnetic measurements; magnetically hard material; microwave devices; temperature coefficient; temperature stability checking; teslameter; type H1-8; Bismuth; Helium; Magnetic materials; Magnetic variables measurement; Magnets; Manufacturing; Microwave devices; Microwave theory and techniques; Stability; Temperature distribution;
Conference_Titel :
Actual Problems of Electron Devices Engineering, International Conference on
Conference_Location :
Saratov
Print_ISBN :
1-4244-0246-8
Electronic_ISBN :
1-4244-0247-6
DOI :
10.1109/APEDE.2006.307440