Title :
10 V measurements with 1V-JVS using a resistive voltage divider
Author :
Real, M. ; Tonina, A. ; Bierzychudek, M. ; Iuzzolino, R.
Author_Institution :
Inst. Nac. de Tecnol. Ind., Buenos Aires, Argentina
Abstract :
This paper describes the design, built and characterization of a resistive voltage divider to compare the 10 V solid state voltage standards output directly with a 1V-Josephson voltage standard. A recent comparison with BIPM exhibits an agreement of 0.024 μV/V with a combined uncertainty uc = 0.04 μV/V.
Keywords :
measurement standards; voltage dividers; voltage measurement; BIPM; JVS; Josephson voltage standard; resistive voltage divider; solid state voltage standards; Contact resistance; Electrical resistance measurement; Electromagnetic measurements; Linearity; Measurement standards; Resistors; Solid state circuits; Thermal conductivity; Thermal resistance; Voltage measurement;
Conference_Titel :
Precision Electromagnetic Measurements (CPEM), 2010 Conference on
Conference_Location :
Daejeon
Print_ISBN :
978-1-4244-6795-2
DOI :
10.1109/CPEM.2010.5545128