Abstract :
We present a 87Rb vapor-cell atomic clock using pulsed optical pumping with an optical detection technique. The pulsed optical pumping technique can eliminate the light shifts and increase the stability of vapor-cell atomic clock. For simplicity, direct laser absorption signal is used for the detection of a clock transition. A frequency stability of about 2×10-11 at 1s has been measured and short term stability is improved by reducing the pulse sequence time with a real time control system.
Keywords :
atomic clocks; frequency stability; optical pumping; rubidium; Rb; clock transition; frequency stability; optical detection technique; pulsed optical pumping; pulsed optically pumped rubidium atomic clock; short term stability; vapor cell atomic clock; Atom optics; Atomic clocks; Laser transitions; Optical detectors; Optical pulses; Optical pumping; Pulse measurements; Pulsed laser deposition; Pump lasers; Stability;