    Verification of performance of commercial LCR meters

  • Author

    Pokatilov, A. ; Satrapinski, A. ; Kübarsepp, T. ; Märtens, O.

  • Author_Institution
    Metrosert Ltd., Tallinn, Estonia
  • fYear
    2010
  • fDate
    13-18 June 2010
  • Firstpage
    408
  • Lastpage
    409
  • Abstract
    Four high-accuracy LCR meters have been compared with the use of Metrosert and MIKES impedance standards. Specifications of LCR meters have been verified at frequency of 1 kHz for the resistance range from 1 Ω to 100 kΩ, for the capacitances in the range from 100 pF to 1 μF and for 100 mH inductor. The performance of LCR meters have been compared also at frequencies from 1 kHz to 1 MHz with ac resistors of different types, at 1 kHz and 100 kHz with 100 pF - 1 μF capacitance standards, and at 1 kHz with 300 μH - 300 mH inductance standards. Results of measurements presented in this summary confirm the specified uncertainty of the tested LCR meters.
  • Keywords
    capacitance measurement; inductance measurement; meters; LCR meter; Metrosert and MIKES impedance standard; inductance capacitance resistance meter; performance verification; resistance 1 ohm to 100 kohm; Calibration; Capacitance measurement; Electrical resistance measurement; Frequency; Impedance; Inductance; Instruments; Measurement standards; Resistors; Testing;