Title :
Application of closed-loop control in the reduction of interference in nerve cuff recordings
Author :
Rahal, Mohamad ; Winter, J. ; Taylor, J. ; Donaldson, N.
Author_Institution :
Univ. Coll. London, UK
Abstract :
A method is presented to reduce the interference pickup in nerve cuff recordings. The gains of differential amplifiers connected to the true tripole nerve cuff arrangement are tuned adaptively to null the residual EMG. Simulation results show that extraction of the neural signal is possible using this method without the need for high order filtering
Keywords :
biocontrol; biomedical electrodes; closed loop systems; differential amplifiers; electromyography; interference suppression; neuromuscular stimulation; adaptive tuning; closed-loop control; differential amplifiers; interference pickup; nerve cuff recordings; neural signal; residual EMG; tripole nerve cuff arrangement; Biomedical engineering; Biomembranes; Differential amplifiers; Electrodes; Electromyography; Impedance; Insulation; Interference; Muscles; Physics;
Conference_Titel :
Electronics, Circuits and Systems, 1999. Proceedings of ICECS '99. The 6th IEEE International Conference on
Conference_Location :
Pafos
Print_ISBN :
0-7803-5682-9
DOI :
10.1109/ICECS.1999.813411