Evaluation of an aperiodic phased array for prostate thermal therapies
Author :
Hutchinson, E.B. ; Hynynen, Kullervo
Author_Institution :
Div. of Health Sci. & Technol., MIT, Cambridge, MA
Volume :
2
fYear :
1995
fDate :
7-10 Nov 1995
Firstpage :
1601
Abstract :
A 57 element aperiodic linear phased array was designed and constructed to investigate the feasibility of using transrectal ultrasound for thermal therapeutic treatment of diseases of the prostate. A method of reducing grating lobe levels by using optimized random distributions of unequally sized elements was used, which made it feasible to use larger elements and hence fewer elements and amplifier channels, while still achieving acceptable power field patterns. The heating capabilities of this array were investigated using an ex vivo perfused kidney for hyperthermia and fresh bovine liver for thermal surgery. The experiments demonstrated that the aperiodic array was capable of generating and controlling adequate power for the purposes of hyperthermia and thermal surgery
Keywords :
biomedical equipment; biomedical ultrasonics; hyperthermia; kidney; liver; radiation therapy; surgery; ultrasonic arrays; amplifier channels; aperiodic phased array; ex vivo perfused kidney; fresh bovine liver; grating lobe levels reduction; medical instrumentation; optimized random distributions; prostate diseases; prostate thermal therapies; therapeutic ultrasound; thermal surgery; thermal therapeutic treatment; transrectal ultrasound; unequally sized elements; Diseases; Distributed amplifiers; Gratings; Heating; Hyperthermia; Optimization methods; Phased arrays; Power amplifiers; Surgery; Ultrasonic imaging;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Ultrasonics Symposium, 1995. Proceedings., 1995 IEEE
Conference_Location :
Seattle, WA
ISSN :
1051-0117
Print_ISBN :
0-7803-2940-6
Type :
conf
DOI :
10.1109/ULTSYM.1995.495861
Filename :
495861
Link To Document :
بازگشت