• DocumentCode
    3492838
  • Title

    Coaxial multiplexer for high accuracy capacitance measurements

  • Author

    Overney, F. ; Corminboeuf, D. ; Moll, E.

  • Author_Institution
    Fed. Office of Metrol. METAS, Bern-Wabern, Switzerland
  • fYear
    2010
  • fDate
    13-18 June 2010
  • Firstpage
    420
  • Lastpage
    421
  • Abstract
    To automate the calibration of capacitance standards, an eight channel coaxial multiplexer has been developed to be used with a widely used commercial capacitance bridge. The use of the multiplexer doesn´t worsen the intrinsic uncertainty of the bridge for capacitors of 10 pF and above. For a 1 pF standard, the use of the multiplexer affect the uncertainty only at frequencies higher than 2 kHz.
  • Keywords
    calibration; capacitance measurement; multiplexing equipment; capacitance 1 pF; capacitance standards calibration; coaxial multiplexer; commercial capacitance bridge; high accuracy capacitance measurements; Bridge circuits; Calibration; Capacitance measurement; Capacitors; Coaxial components; Connectors; Frequency; Measurement standards; Multiplexing; Relays;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements (CPEM), 2010 Conference on
  • Conference_Location
    Daejeon
  • Print_ISBN
    978-1-4244-6795-2
  • Type

    conf

  • DOI
    10.1109/CPEM.2010.5545287
  • Filename