DocumentCode
3492838
Title
Coaxial multiplexer for high accuracy capacitance measurements
Author
Overney, F. ; Corminboeuf, D. ; Moll, E.
Author_Institution
Fed. Office of Metrol. METAS, Bern-Wabern, Switzerland
fYear
2010
fDate
13-18 June 2010
Firstpage
420
Lastpage
421
Abstract
To automate the calibration of capacitance standards, an eight channel coaxial multiplexer has been developed to be used with a widely used commercial capacitance bridge. The use of the multiplexer doesn´t worsen the intrinsic uncertainty of the bridge for capacitors of 10 pF and above. For a 1 pF standard, the use of the multiplexer affect the uncertainty only at frequencies higher than 2 kHz.
Keywords
calibration; capacitance measurement; multiplexing equipment; capacitance 1 pF; capacitance standards calibration; coaxial multiplexer; commercial capacitance bridge; high accuracy capacitance measurements; Bridge circuits; Calibration; Capacitance measurement; Capacitors; Coaxial components; Connectors; Frequency; Measurement standards; Multiplexing; Relays;
fLanguage
English
Publisher
ieee
Conference_Titel
Precision Electromagnetic Measurements (CPEM), 2010 Conference on
Conference_Location
Daejeon
Print_ISBN
978-1-4244-6795-2
Type
conf
DOI
10.1109/CPEM.2010.5545287
Filename
5545287
Link To Document