Title :
A comparative study between different pre-whitening decorrelation based acoustic feedback cancellers
Author :
Essafi, Kawther ; Ben Jebara, Sofia
Author_Institution :
Res. unit TECHTRA, Ecole Super. des Commun. de Tunis, Ariana, Tunisia
Abstract :
The use of an adaptive feedback canceller (AFC) based on signals pre-whitening/filtering in hearing aids seems very attractive, since it limits desired signal degradation when amplification gain is increased. In this paper, we present a comparative assessment of the performances of some existing pre-whitening decorrelation based methods. The used criteria consider the adaptive filter performances, the system stability and the speech quality in terms of distortion and oscillation. Results show that the method including the loudspeaker pre-whitening and the microphone filtering is the best. Moreover, the use of an adequate algorithm for adaptive pre-whitener based on the minimization of a criterion which considers the inter-correlation between the pre-whitened loudspeaker and the filtered microphone signals improves more the performances.
Keywords :
adaptive filters; adaptive signal processing; decorrelation; feedback; hearing aids; loudspeakers; medical signal processing; AFC; acoustic feedback cancellers; adaptive feedback canceller; adaptive filter; adaptive pre-whitener; amplification gain; hearing aids; loudspeaker pre-whitening; microphone filtering; prewhitening decorrelation method; signal degradation; signal filtering; signal prewhitening; speech quality; system stability; Auditory system; Decorrelation; Frequency control; Gain; Hearing aids; Loudspeakers; Microphones;
Conference_Titel :
Multimedia Signal Processing (MMSP), 2010 IEEE International Workshop on
Conference_Location :
Saint Malo
Print_ISBN :
978-1-4244-8110-1
Electronic_ISBN :
978-1-4244-8111-8
DOI :
10.1109/MMSP.2010.5661984