  • Abstract
    A test was conducted of a Hercules Orthoflex capacitive sensor used to measure pressure distribution under the foot for gait analysis and sensory substitution. Maximal nonlinearity for the sensor is 16%, and maximal hysteresis is 10% over the pressure range 0 to 1.3 MPa. Seven transducers in each shoe determine the distribution of pressure.<>
  • Keywords
    biomechanics; biomedical equipment; biomedical measurement; electric sensing devices; pressure transducers; 0 to 1.3 MPa; Hercules Orthoflex capacitive sensor; foot/shoe pressure; gait analysis; maximal hysteresis; pressure distribution; sensor nonlinearity; sensory substitution;