Abstract :
This paper reports on a cone-shaped micro coil for measuring an NMR (Nuclear Magnetic Resonance) signal of an MRI (Magnetic Resonance Imaging). A three dimensional cone-shaped micro coil was fabricated by pulling up an outer frame of a spiral patterned micro coil. Since there are supporting parts between the wiring, the cone-shaped micro coil has uniform pitch. By using this process, we fabricated the cone-shaped micro coil of 30 mm in diameter and 8 mm in height. The resistance and Q-factor of the coil were 2.14 Ω and 32.3 at 8.5 MHz. As a demonstration of MRI by the cone-shaped micro coil, we acquired MR images of a quail egg and an okra.
Keywords :
Q-factor; coils; image resolution; magnetic resonance imaging; microfabrication; nuclear magnetic resonance; 3D cone-shaped microcoil; MRI demonstration; NMR signal measurement; Q-factor; frequency 8.5 MHz; magnetic resonance imaging; nuclear magnetic resonance; okra; quail egg; resistance 2.14 ohm; spiral patterned microcoil; three-dimensional cone-shaped microcoil; Coils; Image resolution; Magnetic resonance imaging; Parasitic capacitance; Resistance; Spirals; Wiring;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Micro Electro Mechanical Systems (MEMS), 2013 IEEE 26th International Conference on
Conference_Location :
Taipei
ISSN :
1084-6999
Print_ISBN :
978-1-4673-5654-1
Type :
conf
DOI :
10.1109/MEMSYS.2013.6474246
Filename :
6474246
Link To Document :
بازگشت