DocumentCode
3497502
Title
Change in strain image contrast with applied compression evaluated via finite elements analysis
Author
Pavan, T.Z. ; Carneiro, A.A.O.
Author_Institution
Dept. de Fis. e Mat., Univ. de Sao Paulo, Sao Paulo, Brazil
fYear
2011
fDate
March 28 2011-April 1 2011
Firstpage
215
Lastpage
215
Abstract
Finite elements analysis (FEA) simulations showed that small strain shear modulus and the nonlinear behavior play roles in the changes of image contrast as the material deforms. The clinical relevance of this work is that nonlinear elasticity of the object being imaged may not only affect the image quality during strain imaging but might also provide information that can be used to differentiate tissues based on their nonlinear behaviors.
Keywords
compressibility; finite element analysis; medical image processing; shear modulus; biological tissues; compression; finite element analysis simulations; image quality; nonlinear elasticity; strain image contrast; Breast tissue; Finite element methods; Imaging; Materials; Strain; Stress; Elastography; Finite Element Analysis; Large Deformation; Nonlinear Elasticity;
fLanguage
English
Publisher
ieee
Conference_Titel
Health Care Exchanges (PAHCE), 2011 Pan American
Conference_Location
Rio de Janeiro
Print_ISBN
978-1-61284-915-7
Type
conf
DOI
10.1109/PAHCE.2011.5871885
Filename
5871885