• DocumentCode
    3497692
  • Title

    Rotational motion effect on sensitivity matrix of MEMS three-axis accelerometer for realization of concurrent calibration using vibration table

  • Author

    Nakano, Atsuki ; Hirai, Yuki ; Sugano, K. ; Tsuchiya, Takao ; Tabata, Osamu ; Umeda, A.

  • Author_Institution
    Dept. of Micro Eng., Kyoto Univ., Kyoto, Japan
  • fYear
    2013
  • fDate
    20-24 Jan. 2013
  • Firstpage
    645
  • Lastpage
    648
  • Abstract
    The rotational motion effect on the sensitivity matrix of a MEMS three-axis accelerometer was evaluated to realize the concurrent calibration of multiple MEMS accelerometers using a multi-axis shaker. The stage position dependency on the sensitivity matrix of a single-mass capacitive SOI three-axis accelerometer was measured. We found that the rotational motions generated by the resonance of the shaker causes an acceleration error on the accelerometer and can be compensated by measuring it using gyroscopes.
  • Keywords
    accelerometers; gyroscopes; micromechanical devices; sensitivity; silicon-on-insulator; vibrations; MEMS three-axis accelerometer; concurrent calibration; gyroscope; multiaxis shaker; rotational motion effect; sensitivity matrix; single-mass capacitive SOI three-axis accelerometer; vibration table; Acceleration; Accelerometers; Calibration; Oscillators; Rotation measurement; Sensitivity; Vectors;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Micro Electro Mechanical Systems (MEMS), 2013 IEEE 26th International Conference on
  • Conference_Location
    Taipei
  • ISSN
    1084-6999
  • Print_ISBN
    978-1-4673-5654-1
  • Type

    conf

  • DOI
    10.1109/MEMSYS.2013.6474324
  • Filename
    6474324