• DocumentCode
    3498510
  • Title

    Control of gaseous mixture in membrane oxygenators

  • Author

    Lautier, A. ; Dehe, T. ; Gaillard, D.

  • Author_Institution
    INSERM, Creteil, France
  • fYear
    1988
  • fDate
    4-7 Nov. 1988
  • Firstpage
    838
  • Abstract
    The inline determination of blood gas pH, PO/sub 2/, and PCO/sub 2/ by either absorptiometric or fluorescence methods has been used for the calculation of gaseous transfers in membrane oxygenators. A prototype gas mixer using a mass flow controller has been developed that allows an accurate gaseous mixture to be obtained. The correct chemical composition is determined either by the perfusionist or automatically by the data obtained from the blood gas sensors.<>
  • Keywords
    artificial organs; biocontrol; CO/sub 2/ partial pressure; O/sub 2/ partial pressure; absorptiometric method; blood gas pH; correct chemical composition; fluorescence method; gaseous mixture control; inline determination; membrane oxygenators;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 1988. Proceedings of the Annual International Conference of the IEEE
  • Conference_Location
    New Orleans, LA, USA
  • Print_ISBN
    0-7803-0785-2
  • Type

    conf

  • DOI
    10.1109/IEMBS.1988.95098
  • Filename
    95098