A pacemaker sensing algorithm for low amplitude p-wave detection
Author :
Kim, Jungkuk ; Huh, Woong
Author_Institution :
Div. of Electron., Inf. & Commun. Eng., Myongji Univ., Yongin, South Korea
Volume :
2
fYear :
1999
fDate :
36495
Firstpage :
1562
Abstract :
A pacemaker sensing algorithm is devised for p-wave detection in low SNR atrial electrograms. The algorithm uses three parameters, amplitude, slew rate, and p-p interval, and a fuzzy neural network structure is adopted to combine their characteristics. The signal characteristic of the current event is compared to those of three most recent p-waves, and the similarity and dissimilarity between them are used to determine if the event is a p-wave. In addition, the distributions of the characteristics of recent p-waves are considered such that a wider distribution takes a less synaptic weight
Keywords :
electrocardiography; fuzzy neural nets; medical signal detection; pacemakers; amplitude; fuzzy neural network structure; low SNR atrial electrograms; low amplitude p-wave detection; p-p interval; p-wave detection; pacemaker sensing algorithm; signal characteristic; slew rate; synaptic weight; Atrial fibrillation; Band pass filters; Fuzzy neural networks; Hemodynamics; Histograms; Humans; Neural networks; Noise level; Organizing; Pacemakers;
fLanguage :
English