Title :
Evaluation of receive-only single coil and four-element phased array for rodent brain imaging at 11.7 T
Author :
Ip, F.S. ; Wosik, J. ; Nesteruk, K. ; Duong, T.
Author_Institution :
Electr. Eng. Dept., Univ. of Houston, Houston, TX, USA
fDate :
March 30 2011-April 2 2011
Abstract :
To achieve high temporal and spatial resolutions in Magnetic Resonance (MR) imaging, one of the keys is to improve the signal-to-noise ratio (SNR) at the signal receiving stage. We used ultra high-field at 11.7 Tesla and custom surface array engineered to fit the size and geometry of a mouse brain in order to increase the SNR and to aim for very high spatial resolution. The coil array was a receive-only coil. Each element of the array was relatively small to reduce the overall loading. The array is cylindrically curved and mounted right above the skull to achieve optimal SNR. Spatial resolution reaches less than 35 micrometers when a four minutes FISP acquisition protocol was applied.
Keywords :
biomedical MRI; brain; coils; image resolution; FISP acquisition protocol; Magnetic Resonance imaging; four-element phased array; mouse brain; receive-only single coil; rodent brain imaging; spatial resolutions; temporal resolutions; Arrays; Coils; Magnetic resonance imaging; Resonant frequency; Signal to noise ratio; Tuning; 11.7 Tesla; phased array; receive-only; rodent brain; ultra high field;
Conference_Titel :
Biomedical Imaging: From Nano to Macro, 2011 IEEE International Symposium on
Conference_Location :
Chicago, IL
Print_ISBN :
978-1-4244-4127-3
Electronic_ISBN :
1945-7928
DOI :
10.1109/ISBI.2011.5872518