Abstract :
Since November of 1998, we have performed four uncertainty evaluations on NIST F-1, our laser-cooled cesium fountain. The results from the latest evaluation have a statistical uncertainty of 2.6×10-15 and an uncorrected bias of 1.1×10-15. This paper presents a brief overview of the evaluations and their uncertainties
Keywords :
atomic clocks; frequency standards; laser cooling; Cs; NIST; NIST F-1; accuracy; cesium fountain primary frequency standard; laser-cooled cesium fountain; overview; statistical uncertainty; uncertainty evaluations; uncorrected bias; Atom lasers; Atom optics; Atomic beams; Fluorescence; Frequency; Laser beams; NIST; Optical attenuators; Optical beams; Uncertainty;