Title :
New calibration technique for coaxial network analyzer reflection measurements at millimeter-wave frequencies
Author :
Horibe, Masahiro ; Ridler, Nick ; Salter, Martin ; Eio, Christopher
Author_Institution :
Nat. Metrol. Inst. of Japan, Tsukuba, Japan
Abstract :
This paper describes the development of a new calibration technique for Vector Network Analyzers (VNA) operating in coaxial lines at millimeter-wave frequencies. The technique utilizes a new design of primary reference standard - the air-dielectric open-circuit. These standards are mechanically robust and provide a simple route to achieving VNA measurements that are traceable to the International System of units (SI). Examples are given showing the calibration technique realized in the 1.85 mm coaxial line size. Results are presented of reflection coefficient measurements made to 70 GHz in this line size.
Keywords :
calibration; coaxial cables; millimetre wave measurement; network analysers; VNA; air dielectric open circuit; calibration; coaxial lines; frequency 70 GHz; international system of units; millimeter wave frequency; reflection measurements; vector network analyzers; Calibration; Conductors; Connectors; Frequency measurement; Reflection; Standards; Uncertainty; Vector network analysis; calibration and measurement; coaxial lines; millimeter-waves; traceability to national standards;