Title :
A digital meter for measuring dissipation factor
Author :
Mahalik, Nitaigour Premchand
Author_Institution :
Dept. of Electron. & Telecommun. Eng., UCE, Orissa, India
Abstract :
The usual practice of measurement of the dissipation factor, i.e. tanδ uses a schering bridge. This method of determining dissipation factor is however, of low accuracy and has a number of disadvantages. To obtain accurate result, for the true balance, stray capacitance are eliminated by using a Wagner´s earthing device which increases the cost. This paper describes a simple and accurate technique for measurement of dissipation factor using a 8085A microprocessor. The implementation scheme is discussed briefly. A computer algorithm for generating lookup table data has been highlighted. The microprocessor programs have been developed by using an assembler. This technique is found to be the most accurate method provided the supply frequency is kept constant
Keywords :
computerised instrumentation; dielectric loss measurement; digital instrumentation; table lookup; 8085A microprocessor; computer algorithm; dielectric loss angle; digital meter; dissipation factor measurement; lookup table; program assembler; Detectors; Displays; EPROM; Microprocessors; Power measurement; Power system measurements; Power system reliability; Reactive power; Time measurement; Voltage;
Conference_Titel :
Properties and Applications of Dielectric Materials, 1997., Proceedings of the 5th International Conference on
Conference_Location :
Seoul
Print_ISBN :
0-7803-2651-2
DOI :
10.1109/ICPADM.1997.616647