DocumentCode :
351848
Title :
Design, simulation and testing of large area silicon drift detectors and detector array for X-ray spectroscopy
Author :
Zhang, W.C. ; Li, Z. ; Siddons, D.P. ; Huang, W. ; Zhao, L.J. ; Li, C.J. ; Kakuno, E.M. ; Pietraski, P.
Author_Institution :
Brookhaven Nat. Lab., Upton, NY, USA
Volume :
1
fYear :
1999
fDate :
1999
Firstpage :
322
Abstract :
Large area (25 mm2) silicon drift detectors and detector arrays (5×5) have been designed, simulated, and fabricated for X-ray spectroscopy. On the anode side, the hexagonal drift detector was designed with self-biasing spiral cathode rings (p +) of fixed resistance between rings and with a grounded guard anode to separate surface current from the anode current. Two designs have been used for the P-side: symmetric self-biasing spiral cathode rings (p+) and a uniform backside p+ implant. Only 3 to 5 electrodes are needed to bias the detector plus an anode for signal collection. With graded electrical potential, a sub-nanoampere anode current, and a very small anode capacitance, an initial FWHM of 1.3 keV, without optimization of all parameters, has been obtained for 5.9 keV Fe55 X-ray at RT using a uniform backside detector
Keywords :
X-ray detection; X-ray spectrometers; silicon radiation detectors; Si; X-ray spectroscopy; detector array; grounded guard anode; hexagonal drift detector; large area Si drift detectors; self-biasing spiral cathode rings; signal collection; sub-nanoampere anode current; symmetric self-biasing spiral cathode rings; uniform backside detector; uniform backside p+ implant; Anodes; Cathodes; Sensor arrays; Silicon; Spectroscopy; Spirals; Surface resistance; Testing; X-ray detection; X-ray detectors;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium, 1999. Conference Record. 1999 IEEE
Conference_Location :
Seattle, WA
ISSN :
1082-3654
Print_ISBN :
0-7803-5696-9
Type :
conf
DOI :
10.1109/NSSMIC.1999.842501
Filename :
842501
Link To Document :
بازگشت