• DocumentCode
    3519248
  • Title

    Time discrimination techniques using artificial neural networks for positron emission tomography

  • Author

    Leroux, J.-D. ; Rouleau, D. ; Pepin, C.M. ; Michaud, J.-B. ; Cadorette, J. ; Fontaine, R. ; Lecomte, R.

  • Author_Institution
    Dept. of Electr. Eng. & Comput. Sci., Sherbrooke Univ., Que., Canada
  • Volume
    4
  • fYear
    2004
  • fDate
    16-22 Oct. 2004
  • Firstpage
    2301
  • Abstract
    Relevant information in positron emission tomography (PET) is currently being obtained mostly by analog signal processing methods. New digital PET scanner architectures are now becoming available, which offer greater flexibility and easier reconfiguration capability as compared to previous PET designs. Moreover, new strategies can be devised to extract more information with better accuracy from the digitized detector signals. Trained artificial neural networks (ANN) have been investigated to improve coincidence timing resolution with different types of APD-based detectors. The signal at the output of a charge sensitive preamplifier was digitized with an off-the-shelf, free-running 100-MHz, 8-bit ADC and time discrimination was performed with ANNs implemented in field programmable gate arrays (FPGA). Results show that ANNs can be particularly efficient with slow and low light output scintillators like BGO (τ=300 ns), but less so with faster luminous crystals such as LSO (τ=40 ns). In reference to a fast PMT-plastic detector, a time resolution of 6.5 ns was achieved with a BGO-APD detector, as compare to 12.7 ns with conventional analog methods using a constant fraction discriminator. With LSO, the ANN was found to be competitive with other digital techniques developed in previous works. In conclusion, ANNs implemented in FPGAs provide a fast and flexible circuit that can be easily reconfigured to accommodate various detectors under different signal/noise conditions.
  • Keywords
    field programmable gate arrays; medical signal processing; neural nets; nuclear electronics; positron emission tomography; solid scintillation detectors; BGO-APD detector; LSO detector; PMT-plastic detector; analog signal processing methods; artificial neural networks; digital PET scanner architectures; field programmable gate arrays; luminous crystals; positron emission tomography; scintillators; time discrimination techniques; timing resolution; Artificial neural networks; Data mining; Detectors; Field programmable gate arrays; Positron emission tomography; Preamplifiers; Signal detection; Signal processing; Signal resolution; Timing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2004 IEEE
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-8700-7
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2004.1462719
  • Filename
    1462719