• DocumentCode
    3520075
  • Title

    A haptic surface scanning and machining parallel manipulator for registration-free bone resurfacing during arthroplasty

  • Author

    Gertler, I. ; Shapiro, Y. ; Wolf, Alon

  • Author_Institution
    Technion - Israel Inst. of Technol., Haifa, Israel
  • fYear
    2013
  • fDate
    6-10 May 2013
  • Firstpage
    2347
  • Lastpage
    2352
  • Abstract
    As orthopaedic implants become more sophisticated and complex, shaping the bone surface to match the implant becomes increasingly important. Accurate bone machining contributes to success of the implant and can potentially increase the implant´s endurance and lifetime. Bone-mounted robots can eliminate intra-operative tracking errors during bone machining. However, registration errors still exist when aligning the pre-operative plan with the patient´s anatomy because the plan is based on pre-operative images. Thus, a robotic system that is capable of executing both surface acquisition and bone machining in a single procedure and within the same coordinate system is preferable to the current state of the art. In this study we used a mini bone-attached robotic system (MBARS) with a haptic interface, to examine the accuracy of force-controlled surface acquisition of a femur model. The average distance error between acquired points and the surface model was less than 1 mm, which was further reduced by 5% with a compensation method which included curvature estimation of the reconstructed surface. An initial evaluation of in-situ bone machining based on these force-controlled scans yielded promising results.
  • Keywords
    data acquisition; haptic interfaces; machining; manipulators; medical robotics; prosthetics; MBARS; arthroplasty; bone machining; bone surface; bone-mounted robots; compensation method; curvature estimation; distance error; force-controlled surface acquisition; haptic interface; haptic surface scanning parallel manipulator; intra-operative tracking errors; machining parallel manipulator; mini bone-attached robotic system; orthopaedic implants; patient anatomy; registration errors; registration-free bone resurfacing; surface acquisition; Accuracy; Bones; Haptic interfaces; Milling; Robot kinematics;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Robotics and Automation (ICRA), 2013 IEEE International Conference on
  • Conference_Location
    Karlsruhe
  • ISSN
    1050-4729
  • Print_ISBN
    978-1-4673-5641-1
  • Type

    conf

  • DOI
    10.1109/ICRA.2013.6630895
  • Filename
    6630895