• DocumentCode
    3526359
  • Title

    Evaluation of a standardization method for 22Na sealed point sources for various measurement conditions

  • Author

    Sato, Yasushi ; Murayama, Hideo ; Oda, Keiichi ; Nishikido, Fumihiko ; Yoshida, Eiji ; Sato, Tomohiko ; Hasegawa, Tomoyuki ; Inadama, Naoko ; Yamaya, Taiga ; Yamada, Takahiro ; Unno, Yasuhiro ; Yunoki, Akira

  • Author_Institution
    Nat. Inst. of Adv. Ind. Sci. & Technol., Nat. Metrol. Inst. of Japan, Tsukuba, Japan
  • fYear
    2010
  • fDate
    Oct. 30 2010-Nov. 6 2010
  • Firstpage
    427
  • Lastpage
    430
  • Abstract
    We have devised a standardization method for a sealed point source that can be applied to nuclides such as 22Na that emit a positron and at least one gamma ray. This method is based on coincidence counting of annihilation radiation and gamma rays using a scintillation detector array. In the present study, we performed Monte Carlo simulations to examine this method for various measurement conditions. EGS5 Monte Carlo simulation results confirm that this method is effective for measurements of a 22Na sealed point source using BGO scintillation detector arrays.
  • Keywords
    Monte Carlo methods; gamma-ray detection; solid scintillation detectors; standardisation; 22Na sealed point source; BGO scintillation detector arrays; EGS5 Monte Carlo simulation; annihilation radiation; scintillation detector array; standardization method; Arrays; Detectors; Materials; Monte Carlo methods; Photonics; Robustness; Standardization;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record (NSS/MIC), 2010 IEEE
  • Conference_Location
    Knoxville, TN
  • ISSN
    1095-7863
  • Print_ISBN
    978-1-4244-9106-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2010.5873794
  • Filename
    5873794