Title :
Micromegas and PIM-like with thermo-bond film frame and spacers
Author :
Liang Guan ; Xiaolian Wang ; HaoHui Tang ; ZiZong Xu ; Tianchi Zhao
Author_Institution :
Univ. of Sci. & Technol. of China, Hefei, China
fDate :
Oct. 30 2010-Nov. 6 2010
Abstract :
Attempts are made to construct micromegas and PIM-like detectors by using thermo-bond film as frames and spaces to support the woven micromeshes and to maintain the gas gaps. The unsupported spans of the meshes in prototypes that we tested are 2.5 cm to 4.5 cm. This method of micromegas construction does not require the photolithographic process. Prototype construction and test results of the single avalanche gap micromegas and the double gap Parallel Ionization Multiplier (PIM) are reported. The best energy resolution obtained for 5.9 keV x-ray is 13.7% (FWHM) for Micromegas prototypes. The test results also show it is possible to use woven thick mesh to build PIM with sufficient gain and large component of fast electron signal.
Keywords :
ionisation chambers; position sensitive particle detectors; PIM-like detectors; double gap parallel ionization multiplier; energy resolution; fast electron signal; gas gaps; micromegas construction; micromegas prototypes; micromeshes; photolithographic process; single avalanche gap micromegas; thermobond film frame; Anodes; Detectors; Electric fields; Energy resolution; Films; Prototypes; Weaving;
Conference_Titel :
Nuclear Science Symposium Conference Record (NSS/MIC), 2010 IEEE
Conference_Location :
Knoxville, TN
Print_ISBN :
978-1-4244-9106-3
DOI :
10.1109/NSSMIC.2010.5874014