DocumentCode :
3530542
Title :
Accurate measurements of the rise and decay times of fast scintillators with solid state photon counters
Author :
Seifert, S. ; Steenbergen, J.H.L. ; van Dam, H.T. ; Vinke, R. ; Dendooven, P. ; Löhner, H. ; Beekman, F.J. ; Dorenbos, P. ; van der Kolk, E. ; Schaart, D.R.
Author_Institution :
Delft Univ. of Technol., Delft, Netherlands
fYear :
2010
fDate :
Oct. 30 2010-Nov. 6 2010
Firstpage :
1736
Lastpage :
1739
Abstract :
In this work we present a measurement setup for the determination of scintillation pulse shapes of fast scintillators. It is based on a time-correlated single photon counting approach that utilizes correlation between 511 keV annihilation photons to produce start and stop signals in two, separate crystals. The measurement is potentially cost effective and simple to set up while maintaining an excellent timing resolution of 120 ps. Apart from these advantages this method probes the entire volume of the crystal under investigation rather than a thin layer at the surface. Here, we present a proof-of-concept and first results for the time constants of LYSO.
Keywords :
photon counting; solid scintillation detectors; LYSO time constants; annihilation photons; decay time; fast scintillators; rise time; scintillation pulse shapes; solid state photon counters; Crystals; Detectors; Histograms; Photonics; Shape; Timing; Fast Scintillator; LYSO; MPPC; SiPM; Solid State Photon Counter; decay time; rise time; time constant;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record (NSS/MIC), 2010 IEEE
Conference_Location :
Knoxville, TN
ISSN :
1095-7863
Print_ISBN :
978-1-4244-9106-3
Type :
conf
DOI :
10.1109/NSSMIC.2010.5874071
Filename :
5874071
Link To Document :
بازگشت