  • Abstract
    The aim of this study was to develop of a flow-through multi-wire proportional counter with high sensitivity for positrons emitted from the commonly used positron emitters 11C, 15O, 18F and 68Ga. The detector was designed for use in studying time-activity curves in rats after i. v. injection of [15O]water.
  • Keywords
    biomedical measurement; blood; blood flow measurement; positron emission tomography; detector; flow-through multiwire proportional counter; on-line blood sampling; positron-emitting radionuclide measurement; time-activity curves; Blood; Calibration; Detectors; Humans; Positron emission tomography; Positrons; Radiation detectors;