• DocumentCode
    3533566
  • Title

    A prototype PET detector module using micro-channel plate photomultiplier tube with waveform sampling

  • Author

    Kim, Heejong ; Kao, Chien-Min ; Chen, Chin-Tu ; Frisch, Henry ; Genat, Jean-Francois ; Tang, Fukun

  • Author_Institution
    Dept. of Radiol., Univ. of Chicago, Chicago, IL, USA
  • fYear
    2010
  • fDate
    Oct. 30 2010-Nov. 6 2010
  • Firstpage
    2750
  • Lastpage
    2753
  • Abstract
    We are developing a prototype PET detector module as an effort to demonstrate and investigate the feasibility of using large area flat panel (e.g. 8"×8") micro-channel plate photo-multiplier (MCP PMT) for PET instrumentation. The detector module consists of an array of LSO (or LYSO) crystals coupled to a commercially available MCP PMT (Photonis XP85022) with transmission-line strips and read-out by a waveform sampling electronics employing the DRS4 chips at the end of the strips. The details of building and testing of the prototype PET detector module is presented and the benefits of using waveform sampling in PET application also will be discussed.
  • Keywords
    lutetium compounds; photomultipliers; positron emission tomography; readout electronics; signal sampling; transmission lines; yttrium compounds; (LuY)2SiO5; DRS4 chips; LuSiO; MCP PMT; microchannel plate photomultiplier tube; prototype PET detector module; readout; transmission line strips; waveform sampling; Detectors; Energy resolution; Positron emission tomography; Prototypes; Timing; Transmission lines;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record (NSS/MIC), 2010 IEEE
  • Conference_Location
    Knoxville, TN
  • ISSN
    1095-7863
  • Print_ISBN
    978-1-4244-9106-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2010.5874293
  • Filename
    5874293